telescopeѲptics.net
▪
▪
▪
▪
▪▪▪▪
▪
▪
▪
▪
▪
▪
▪
▪
▪ CONTENTS
◄
12. TELESCOPE EYEPIECE
▐
12.3. Eyepiece aberrations II
► 12.2. Eyepiece aberrations I : spherical, coma, astigmatism, field curvatureWhile the primary goal of the telescope objective is to transform flat incoming wavefronts into spherical, the eyepiece needs to accomplish exactly the opposite. If the wavefront entering the eye is not flat, the point-image on the retina will suffer from wider diffraction energy spread, just as it does when the wavefront formed by the objective deviates from spherical. The extent to which it will affect perceived image quality depends on the detail size and properties, as well as image magnification. If the front eyepiece focal plane doesn't coincide with the image plane of the objective, the wavefronts originating from object-image points will not exit the eyepiece flat, but curved (nearly spherical) and, as such, will be rejected by the eye in favor of the flat wavefronts coming from a point coinciding with the front focal plane of the eyepiece. The resulting error is defocus, easily correctable by moving the eyepiece to the proper position. If the object-image and the front focal plane coincide, and wavefronts entering the eyepiece are spherical, it will emerge from it flat if ocular is aberration-free. In the real world, these exiting wavefronts will be aberrated to some extent, and the aberrations are spherical, coma, astigmatism and field curvature, chromatism, image distortion and spherical aberration of the exit pupil. Due to a number of lens elements, aberration expressions for eyepieces are lengthier than for the objective. Also, eyepiece specs are commonly not known. For those reasons, eyepiece aberrations will be considered only in general terms. Before addressing specific aberrations, here's illustration of the main parameters of eyepiece aberrations using a simple Ramsden-type configuration.
While even a single lens eyepiece can produce a good axial correction - due to the generally small cone footprint on eyepiece surface - its field correction is very poor, with the off-axis aberrations increasing exponentially due to the larger footprint combined with stronger radii. For instance, the eyepiece shown is a 10mm f.l. Ramsden with a pair of BK7 PCX lenses w/8.5mm radius of curvature, 6mm lens separation and 5.7mm from field lens to objective's image, as well as from eye lens to exit pupil. At f/10, it is aplanatic, flat field with low astigmatism (0.11 wave RMS 15° off axis). If the eye lens is taken out, both, footprint and refraction (requiring r.o.c. reduction to 5.2mm) at the field lens - which already contributes nearly all off the astigmatism - have to increase, resulting in several times larger astigmatism severely limiting usable field. In addition, multiple elements may allow for partial offsetting of aberrations between them, further reducing their final level. For that reason, eyepieces utilize two or more lenses. In the above example, the field lens "sees" the image formed by the objective as its object. Its aperture stop, however, is at the objective, because it is where the chief incident off axis ray intersects the axis. The eye lens "sees" the virtual image, formed by reverse extension of rays refracted by the field lens, as its object, and its aperture stop is at the point of intersection of the extended chief raj converging from the field lens and the axis. Thus, to both apply the lens aberration relations for displaced aperture stop. Obviously, in order for the eye lens to produce a collimated pencil, its object - the image formed by the field lens - has to be at the distance equaling its focal length. In this particular example, specific aberrations are very low spherical aberration, less than 1/100 wave P-V, most of it at the second lens due to the wider footprint (cone aperture) on it. Coma is practically zero on both lenses, due to the specific combination of object distance and stop separation for each lens. If, for instance, the focal length of lens changes, it changes the object and aperture stop locations relative to the lens' focus point, and the aberration is reintroduced, rather quickly. Nearly all of the astigmatism is on the field lens, and it is of opposite sign to that on the eye lens. The astigmatism is of the opposite sign to the Petzval curvature, offsetting it to produce flat, mildly astigmatic image surface. Visual detection of aberrations in the eyepiece doesn't depend on their magnitude, but on their apparent angular size. This applies to the aberrations that are enlarging central diffraction maxima, like coma and astigmatism; spherical aberration actually makes the central maxima smaller as it increases in magnitude, manifasting only as a loss of sharpness and resolution. The effect of off ais aberration will become just visible to the average eye as it reaches about 3 arc minutes in size. That is when point-like source start appearing as extended, although their shape is being clearly recognized at about 5 arc minutes apparent size (these numbers vary individually, possibly significantly). Knowing that the Airy disc diameter in eyepiece is given by 4.6F/f arc minutes (for 550nm wavelength), F being the telescope focal ratio and f the eyepiece focal length, the corresponding eyepiece focal lengths for recognizing the Airy disc are 1.5F and 1F mm, respectively. Since we don't see the Airy disc, but the central maxima which is always smaller, the actual focal lengths are also somewhat smaller. With the astigmatic blur close to field edge being up to several times the Airy disc diameter, or more, particulary in the conventional eyepieces with fast focal ratio systems, it can be detected even at the lowest magnifications. Eyepiece spherical aberrationThe wavefront error of spherical aberration in the eyepiece changes with the fourth power of the objective focal ratio. Thus, halving the objective F-number increases the wavefront error by a factor of 16. This makes eyepiece correction for spherical aberration critical. Even eyepieces superbly corrected at mid-to-small focal ratios, may become noticeably affected at large relative apertures. Since spherical aberration in eyepieces is also in proportion with the eyepiece focal length, longer f.l. units are more affected. However, they normally don't produce sufficient magnification to show the effect of aberration. Degree of eyepiece correction for spherical aberration also varies with the eyepiece type and brand. Best corrected conventional 5mm f.l. eyepieces will have ~ 1/10 wave P-V of spherical aberration (under-correction) at f/4, while those on the opposite end are closer to 1/4 wave P-V. This doesn't necessarily mean that the telescope performance will be noticeably affected. Most any objective contributes certain amount of spherical aberration of its own. If it is of the opposite sign to that of the eyepiece, the final result may still be admirable. But if the error is of the same sign for both, it will result in inferior performance. For instance, if the objective is 1/5 wave under-corrected, and the eyepiece 1/8 wave under-corrected, their cumulative error will be 1/3 wave P-V of under-correction. That is another example of the actual telescope aberration level being determined by the combined effect of both, objective and eyepiece.
Modern-type eyepieces employing Smyth lens (negative field lens)
behave differently. In general, they tend to induce over-correction,
which also may become significant at very fast focal ratios, especially
when combined with the correction error of the objective that is of the
same sign. The Nagler-type eyepiece (Type 1) presented in the Rutten-Venrooij's
"Telescope Optics" raytraces to nearly 1 wave P-V of over-correction at
f/5. Since it is a 100mm f.l. unit, the error would drop to nearly 1/20
wave P-V with a 5mm f.l. unit at f/5. At f/4, it would be nearly
1/8 wave P-V. Should the objective be 1/6 wave P-V over-corrected at
f/4, the combined error would come to ~1/3.5 wave P-V of
over-correction.
As an illustration of eyepiece performance at very fast focal ratios, here's
raytrace of three actual designs: Nagler's patent from 1988(b), exceptionally
well corrected for spherical aberration even among the modern, complex designs (top),
Nagler's Plossl patent (bottom left), and a modern monocentric. The ray spot
plots are given for field center (left) and outer field point given below
focal length value (right).
Going from f/5 to f/3, this Nagler 2 does generate several times more
of spherical aberration, but since the error at f/5 is so small - the
Strehl indicates 0.003 wave RMS, or 1/100 wave P-V - it only gets to
0.022 wave RMS at f/3. The Plossl, which is near the top correction-wise
for the conventional eyepieces, already falls short at f/3. However, the
modern monocentric still holds it's ground with 0.9+ Strehl. Astigmatism
changes in inverse proportion to the square of f-ratio.
Eyepiece comaOff-axis aberrations are more pronounced in eyepieces, due to their large viewing angles. Coma is, in general, not significant in eyepieces due to it being usually minimized by design, and much lower than astigmatism, which cannot be reduced nearly as efficiently. The amount of coma wavefront error in a positive lens changes in inverse proportion to the square of the focal length, and in proportion to the third power of the cone width. In effect, for given field angle, it changes in proportion to the eyepiece focal length. It also changes in proportion to the third power of telescope focal ratio. Given eyepiece design, the 30mm f.l. unit will have three times the coma of the 10mm f.l. unit. Either will have eight times more of the coma wavefront error at f/5, than at f/10 telescope focal ratio.
As with the coma
originating
at the objective, eyepiece coma increases with the off-axis height in
the image plane.
However, it is
eyepiece astigmatism that usually dominates. It also depends on the lens focal
length and the cone width, being inversely proportional to the eyepiece f.l. and in proportion to the square of the cone width. In effect, for
given field angle it
is, as coma, proportional to the eyepiece f.l. but, unlike coma, it
changes in proportion to the square (not third power) of the telescope
focal ratio (note that the geometric blur size changes in proportion to
the focal ratio, but also with the square relative to the Airy disc). In other words, the wavefront error of astigmatism is four
times larger at f/5 than at f/10. Unlike coma, eyepiece astigmatism
increases with the square of apparent angle.
Also, unlike either spherical aberration or coma, lens astigmatism
- with the stop at the surface - is
independent of all three - refractive index, lens position factor and
shape factor; it only depends on the lens focal length. The means of controlling it with a set of positive lenses and lens
groups is generally limited, but the main reason why is it typically
strong in conventional eyepieces is that it is the third at the list of
priorities, after spherical aberration and coma. It is significantly
more complicated - or impossible, with simple designs - to have all three corrected.
0.01fe(ε/F)2<W<0.03
fe(ε/F)2,
in units of 550nm wavelength, where
fe
is the eyepiece f.l., ε the eyepiece apparent field angle in degrees
(radius),
and F, as before, the telescope focal ratio. Plots at left show
the range of astigmatism based on this approximation for three telescope
ratios, fast, medium and slow. It scales with the eyepiece focal length.
Taking 20-degree
near-edge AFOV for conventional eyepieces, it gives the following P-V
values of astigmatism for selected focal ratios and eyepiece focal
lengths:
The error scales with the square of AFOV, which means
it is four times smaller at 10° off axis, and 16
times smaller at 5°. The lower limit is not
strict; the error can be still lower, but not significantly. On the
other hand, poor designs can have significantly more astigmatism than
the upper limit, up to twofold, or so. It can be assumed, however, that
most of the conventional eyepieces made these days are closer to the
lower limit shown in the table.
For given magnification, eyepiece astigmatism scales
inversely to the focal ratio. For instance, if an �/4
with 10mm f.l. eyepiece has 2.5-7.5 waves P-V of astigmatism at 20
off-axis, an f/8 system with 20mm
eyepiece of identical design will have 1.25-3.8 waves.
An interesting example is Konig design given in "Telescope Optics" from
Rutten and Venrooij. It has practically cancelled primary
astigmatism (and well corrected higher-order term), at a price of somewhat
stronger than usual coma (FIG.
213). In terms of the RMS wavefront error, at 10° off-axis it is -
according to OSLO - superior to its low-coma, strong astigmatism variant
at both, f/10 and
f/5, with 0.033 and 0.23 vs. 0.065 and 0.33 wave,
respectively. At 20° field angle even more so: 0.16 and 0.7 vs. 0.94 and
3.7 wave. The enormous increase in the RMS error in the astigmatic Konig
variant reflects the devastating effect of combined lower- and
higher-order astigmatism. Yet, it is the type more likely to be found on
the market, because of the general notion that coma is less desirable
aberration form.
What actually puts the limit to usable field in a typical conventional
eyepiece is the combination of lower- and higher-order astigmatism. At
relatively small field angles, usually below ~15°, the higher-order
component is, in most properly designed eyepieces, negligible to
non-existent. But at once it creeps in at larger angles, it quickly
explodes with the 4th power of field angle, adding to the already
large lower-order component. This puts an end to the acceptable field
size.
As mentioned, eyepiece astigmatism
diminishes with the eyepiece focal length. A 10mm conventional eyepiece
unit has ~3 times lower astigmatism than 30mm unit, resulting in about
1.7 times larger linear diffraction limited field. However, the
difference is not that obvious in the eyepiece, due to shorter f.l.
eyepieces having proportionally larger magnification, which mainly
offsets the aberration decrease (FIG. 210).
FIGURE
210: Off-axis aberration in a conventional eyepiece (from
~10" distance the view is similar to the actual one). The difference
in the nominal level of aberration - from 16 waves of
astigmatism combined with 2.4 waves of coma for the 30mm f.l. at
f/5, to 0.65 wave of astigmatism and negligible coma
for the 5mm at f/10, both at 20° off-axis, which is more or less the
typical level - is much greater than
what the actual view shows. Actual Airy discs are not
of identical size at these three f-ratios (diameter being given by 4.6F/�e
in arc minutes), for 550nm wavelength; it would be
noticed visually, assuming sufficient magnification, but not with
the given screen resolution. Note that the plot shows only
aberrations originating in the eyepiece; the final spot shape and
size vary with the amount of off-axis aberration induced by the
objective, as well as with the combined field curvature of the
objective and eyepiece. In general, eyepiece aberrations -
particularly astigmatism - dominate in fast to medium telescope
systems.
For meaningful correction of eyepiece astigmatism it is necessary to
introduce a negative (Smyth) lens, which induces astigmatism of
opposite sign to that of the positive lens group. It also induces
Petzval field curvature of the opposite sign, resulting in both
astigmatism and field curvature minimized. Best known brand of this
kind, the Nagler, has astigmatism reduced up to several times vs.
comparable conventional eyepieces.
Field curvature in the telescope eyepiece is directly related to its
astigmatism. A hypothetical astigmatism-free eyepiece would form the
image coinciding with the Petzval surface. Being formed by a positive
power lens system, this surface is concave toward the eyepiece and,
considering relatively short focal lengths, rather strongly curved
(eyepiece's Petzval curvature, as well as other aberrations, are
determined by reverse ray trace, with collimated light entering through
the eyepiece's exit pupil).
Real world eyepieces, however, produce strong astigmatism,
particularly the conventional types. Usually, this astigmatism is of
opposite sign to the Petzval, thus abaxial points form astigmatic
surfaces less curved relative to the Petzval up to a certain level. At
the point when the sagittal astigmatic surface is half as curved as
Petzval curvature, best image surface is nearly flat (FIG. 211B).
Further increase in the astigmatism causes best image surface to become
increasingly curved to the opposite side (FIG. 211C).
It is possible that
astigmatism in the eyepiece is of the same sign as its Petzval (for
instance, due to unbalanced higher-order astigmatism) in which case the astigmatic image surfaces are closer to the eyepiece
than Petzval's, thus more strongly curved, and more so as
the astigmatism increases.
FIGURE 211: Astigmatism-free eyepiece will form flat field if the image formed by the objective is
also free from astigmatism and coincides with the eyepiece's
Petzval surface P
(A).
Astigmatism modifies the field curvature depending on its sign and
magnitude (B,C). If the sign is opposite to the Petzval's - usually the case -
sagittal astigmatic surface S forms on the convex side of
the Petzval's. Tangential surface T is always 3 times farther away from
the Petzval than sagittal, with the best, or median surface M
midway between the two.
Note that the curves show primary astigmatism. Eyepieces with strongly
curved lenses also generate secondary, higher-order astigmatism. Since it has significantly different rate of
increase than the lower-order form (4th and 2nd power of the field
radius, respectively), once it reaches considerable level at larger
field angles, it also strongly alters the initial field curvature.
Depending on the design particulars, the effect with respect to field
curvature can be either positive
(field flattening) or negative; it can also be positive up to a certain
image radius, then negative, or vice versa.
Flat eyepiece field means that all off-axis pencils exiti the eye lens
collimated, enabling the eye to focus on all points across the field
simultaneously. When the eyepiece field is curved, off axis pencils are
progressively more diverging, or converging, going farther off axis,
hence they cannot be focused simultaneously with the central pencils on
the retina (i.e. image points appear to be closer, for diverging
pencils, or farther away, for converging ones). This can be either
lessened or worsened when combined with the curvature of the image
formed by the objective, as shown below.
For simplicity, only best image surface is shown, i.e. field curvature
is assumed to be either the Petzval with zero astigmatism, or best
(median) astigmatic surface, when astigmatism is present. This is
simplification in that it is not only the image curvatures, but also
their respective astigmatism that interact, and since astigmatism
directly influences field curvature it is a factor that shouldn't be
entirely neglected. However, with the astigmatism of the objective being
typically negligible in comparison, the astigmatism factor can be
neglected in this respect for a general consideration, and assume that
the interactions of field curvatures alone will give a good
approximation of the combined field curvature.
Most
objectives generate curvature concave toward objective, and most
eyepieces nowadays have near flat field, in which case the combined
visual field has curvature similar to that of the objective (1a).
The reason is that only light from those points of the objective's image
laying in the front focal plane of the eyepiece will exit the eyepiece
as collimated pencils. Those farther out will exit converging, hence eye
will focus on the (defocused) point of the objective's image laying in
the front focal plane of the eyepiece. In order to focus on the points
curving out away from the eyepiece, the eyepiece has to be shifted
toward them. In effect, the visual image field is curved away from the
eye. Eyepiece itself can generate curved image surface, either convex or
concave toward the eye. If it is convex and nearly coinciding with the
image surface of the objective, light from all points will exit eye lens
as collimated pencils, i.e. will focus onto the retina, and the visual
field will be effectively flat (1b). If, however, eyepiece field
curvature is opposite to that of the objective's, the effective visual
field curvature will be a sum of the two, twice as strong (1c).
Likewise, if the objective's image field is flat, the visual field will
have the curvature of the eyepiece's field (2a-c). With some
objectives which generate field curvature convex toward them, like
Gregorian two-mirror telescope, the combined field curvature forms in
the same manner, only the net sum with the same eyepieces is different (3a-c).
Astigmatism and field curvature of the eyepiece combine with those of
the objective, to form astigmatism and field curvature of the final
visual image. Just as the hypothetical astigmatism-free eyepiece would
need objective's image to be astigmatism-free, with the two Petzval
surfaces coinciding, in order to produce flat, astigmatism-free combined
field, an astigmatic eyepiece would need objective whose astigmatic
image would coincide with its own to result in astigmatism-free combined
image. Eyepiece astigmatism is normally significantly stronger than that
of the objective, especially for longer f.l. conventional eyepieces.
Hence the astigmatism of the objective doesn't have much of effect on
the final image: it is mainly determined by the eyepiece.
Short focal length eyepieces have proportionally lower astigmatism (as the transverse aberration; lower to the square of it as wavefront error, due to the smaller Airy disc), and
it can be more noticeably affected by the astigmatism of the objective.
Astigmatic surface profile of most amateur telescopes (Newtonian,
refractor, Cassegrain) is roughly similar in form to that shown on
FIG. 211C, concave toward objective, with the tangential
surface closer to it. Newtonian form is identical to it, while in
refractors and Cassegrain-like systems - including SCT and Gregory MCT -
sagittal surface is also concave toward converging light, and so is the
system Petzval; in Maksutov-Cassegrain with separate secondary, system
astigmatism can be nearly corrected, with the field curvature nearly
coinciding with its Petzval.
If the eyepiece has, say, form of astigmatism shown in (B),
twice stronger than the
astigmatism of the objective (i.e. double the sagittal-to-tangential
surface separation) having the form shown in (C), the tangential surface of the combined image will
be flat, with the sagittal unchanged, for the combined astigmatism
half that in the eyepiece alone, and median image curvature of opposite
sign, but of the same magnitude as that of the objective (Petzval surface becomes a factor only in
astigmatism-free systems). If the eyepiece has the same amount of
astigmatism, but with tangential surface flat and sagittal twice as
strong as in the former example (which requires stronger eyepiece
Petzval curvature), tangential surface in the combined image will have
curvature equal to that in the objective, with the sagittal surface
being as curved as in the eyepiece; the combined astigmatism is 50%
greater than in the eyepiece, but with the combined median surface
curvature somewhat weaker.
Gregorian-like two-mirror systems have, like Newtonian, Petzval surface
curvature of the same sign as typical eyepiece's Petzval, but considerably
stronger, with both sagittal and tangential astigmatic surface usually
convex toward converging light. With both telescope types, this allows
for the possibility of correcting both astigmatism and Petzval surface
curvature in the final image with matching eyepiece (easier in the Gregorian, whose Petzval
curvature is significantly closer in magnitude to that of the eyepiece).
Note that strong field curvature in eyepieces, if mainly result of
astigmatism as shown on FIG. 211C, does not implicate significant
defocus error, even without the ability of the eye to accommodate.
Sagittal surface is usually relatively weak, and the wavefront error
along it is identical to that along the tangential surface, with the
wavefront error along best (median) surface smaller by a factor of
1/√1.5.
In other words, it is the error of astigmatism that dominates
(but if the eye is unable to accommodate to the best image curvature,
the point images will be elongated).
Needed eye accommodation to eyepiece field curvature (calculated based
on the thin lens Gaussian equation, thus only approximation)
depends on three factors, linear defocus of the off axis point in the
visual focal plane(∆), eyepiece f.l. (f) end eye f.l. (E) as:
A=(∆E A=∆ED/[f
for that defocus i.e. needed accommodation in diopters,
where D is the eye focal length in diopters (D~59; while the
optical f.l. of the eye, due to the image medium (n~1.33) is about 17mm,
giving D=1000/17~59, its physical f.l. is about 23mm). Most often,
point defocus ∆ is sufficiently small vs. eyepiece focal length,
and the relation simplifies to A~1000∆/f2. Taking A=1,
for defocus corresponding to one diopter, gives ∆~f2/1000.
Or, put simply, light from a point displaced
from the point-image defined surface in its image space
(which is in the actual setup the objective's image space, and object space
for the eyepiece), will not form a parallel exit pencil, but either slightly
converging one - for a point farther away - or diverging, if closer than the
distance producing collimated exit pencil.
For instance, a field edge point 0.1mm off that imaginary point-surface in a 10mm f.l.
eyepiece will be as much removed from the actual (objective's) image point in the
flat image plane of the objective, thus the actual point it reimages will not form a
parallel exit pencil, but slightly converging (actual field point farther away,
illustration at left), or diverging (actual field point closer). Since change in the focal
length f of xf approximately corresponds to bringing object from infinity to a distance
of f/x (for x~0.1 and smaller), 1% difference in point distance vs. f.l. of the eyepiece means
that the rays coming to the exit pupil won't be parallel, but either converging, as if coming
from a point 100f=1m away, or diverging, as if coming from an imaginary point at the same
distance but behind the objective (eyepiece). Since 1m vs. eye focal length numerically
represents the eye focal length in diopters (59D), the defocus vs. infinity will be
approximately 1/59 of the focal length, or 1 diopter. For the case shown, with the actual
point farther away from the point that would form a parallel exit pencil, exiting rays will
be converging, forming in the relaxed (infinity) eye focus point shorter than
infinity focus, hence requiring eye lens relaxation (streaching) beyond that needed for
infinity in order to bring it to the retina.
Despite being numerically identical, only positive for extending, and
negative for shortening the focus, the two forms of accommodation are
very different to the eye. Shortening the focus, which requires
compressing the eye lens to a stronger radii than in its infinity
mode, is natural to the eye, and much easier than streching eye lens
out to weaken the radii, which it can do in a much more limited way.
Thus, assuming flat objective's image field, eyepiece field curvature
convex toward the eye is much preferred to the opposite shape (note
that field curvature resulting from reversed raytracing has opposite
sign of its actual field curvature, which is the case with distortion
as well; in systems w/o reversing reflection, light travels from left
to right, so the proper eyepiece orientation is with exit pupil on the
right side; in a Cassegrain-like system, eyepiece will have the
orientation shown, but the objective's field curvature will be
reversed due to reflection).
Above formulae imply that scaling eyepiece up results in lower
accommodation required, in inverse proportion to the focal length
(not in inverse proportion to the square of it, as it appears,
because the value of Δ also changes, in proportion
to the focal length). Graph below allows to extract the values
for point displacement and accommodation requirement for
eyepieces of different focal lengths. On the horizontal scale
it shows relative point displacement values, accounting for the
change in Δ with eyepiece focal length f, and on
the vertical scale the relative value of corresponding required
accommodation, given as the actual accommodation in diopters multiplied
with f/10 factor (these are a consequence of using Δ
that changes with eyepiece focal length in the above equation
for accommodation).
For instance, the actual point displacement for 0.1mm point
displacement in the 10mm f.l. unit, given by (10/f)Δ,
f being the eyepiece focal length, will be a bit less than
0.05mm in the 5mm unit, and 0.2mm in the 20mm unit. The
corresponding accommodation requirements will be 2 and 0.5 diopters,
respectively (from the value on the vertical scale divided with f/10).
Value of Δ alone represents point displacement
corresponding to 1 diopter accommodation; it is obtained by dividing the
scale value - in this case 1 - with 10/f. For the 5mm unit, it is
slightly less than 0.025. In other words, value of Δ corresponds to the scale value
only for the 10mm unit, as well as the scale value for accommodation requirement.
For different focal lengths, both change as indicated. The plots are
not straight lines due to the form of the equation, with those for
f<17mm being concave, and for f>17mm convex in the point of origin, 0;0
(the larger differential vs. 17mm f.l. the more so).
It is obvious that the same amount of accommodation will be required
if the situation is reversed, i.e. with a flat-field eyepiece, and
objective having the same amount of image curvature. However, since
the radius of curvature is constant, and field sagitta (s) for the
latter is given by s=h |